Quantitative Analysis of Salithion by Gas Chromatography
Bi Fuchun Wang Wenli

Abstract
Reference
Related Articles
Download: PDF (0KB)   HTML 0KB   Export: BibTeX or EndNote (RIS)      Supporting Info
Abstract The quantitative analytical method for salithion by a wide bore capillary column has been studied.The optimum gas chromatographic conditions were as follows.Column:fused silica,HP 1 (cross linked methy1 silicone) 10m��0 53mm��2 65��m film;column temperature:150��;injector and detector temperature:200��;carrier gas:H 2 15mL/min Di n propy1 phthalate was used as the internal standard.The analytical method for salithion gives better accuracy and precision.